Biological Safety Cabinets/Workstations/Enclosures

Here:Homepage > Products > Experimental Containment > Biological Safety Cabinets/Workstations/Enclosures

Class II Biological Safety Cabinet, Type B2

Ideal for biological applications using chemical reagents


DreamLab® DL-BSCB series Class II Biological Safety Cabinets Type B2 provide both operator protection and process/sample protection for many biological applications utilizing hazardous chemicals which can not be removed by HEPA filtration. Unlike that in type A2, airflow in type B2 biological safety cabinets is unidirectional and finally goes out through outlet connecting to general ducting system instead of partially recirculating. 


Airborne contaminants are captured in incoming HEPA filtration before entering the active workspace. The exhaust air is also HEPA filtered so harmful biological aerosols are not spread out of the room. To diminish the threat of cross-contamination, the built-in timed UV light can be used to irradiate work surfaces and tools between samples or cases. Full-closure sash and safety switch prevent operator exposure to UV energy. DLTouch microprocessor controller constantly monitors filter conditions and alerts the user if filter replacement is needed.



DL-BSC150B B2型II级生物安全柜-斜面.jpg

DL-BSC100B Class II Biological Safety Cabinet, Type B2


标准特性图标.pngStandard Features

-  DLTouch automatic safety controller 

-  Rigid epoxy-coated steel construction

-  Single piece type 304 stainless steel interior liner with coved corners for ease of cleaning and decontaminating

-  Total exhaust design provides operator and process protection

-  HEPA filters in incoming and outgoing airflow

-  Sliding front sash frame for full chamber access

-  Clear polycarbonate sash for visibility and UV reflection

-  Angular front grill design

-  Ergonomic armrest 

-  Timed shortwave UV light

-  Safety switch to prevent exposure to harmful UV energy

-  Integral LED light

-  Constant temperature and humidity monitoring

-  Lab event timer

-  No ductwork required

-  No installation required, plugs directly into a standard 110V or 220V electrical outlet

-  Extremely quiet operation < 60dBA

-  Shipped fully assembled and tested

-  Available in 110V AC or 220V


附件选项图标.pngOptions

-  Mobile cart with casters

-  Electric control of sash lift

-  ULPA filters


工作原理图标.pngHow does it work?

B2型II级生物安全柜气流图.jpg


  1. Room air enters at A where it is pulled into the incoming HEPA filtration on the top and where it is pulled into the inlet grille along with a portion of the chamber air.

  2. Incoming air passes through prefilter and HEPA filter at B, forming a stable ISO 5 downflow.

  3. Chamber air is mixed with room air travelling through inlet grille at C before moving to the exhaust filtration area.

  4. Exhaust air moves through the HEPA filter at D and accesses to outlet duct E connecting with room ventilation system. 

  5. Chemical gases in the airflow are handled by remote devices. 



WechatIMG1704.pngModels and Dimensions

DL-BSCB系列型号尺寸.png

更多信息图标.pngMore Information ……

About filtration, click here;

About DLTouch automatic safety controller, click here;

About applications, click here.

 

To download brochure of this product, click here.